Thermo Scientific
GenPure
21
6
The Installation Area
Take the following criteria into consideration when selecting the installation area:
Feedwater pressure, not below 0.1 bar (1.45 PSI) and not above 6 bar (87 PSI).
The feedwater pressure must never exceed 6 bar. If it is higher than this, install an
additional external Pressure regulator.
• Minimum air tempe2°C.
• The surface on which the system is installed must be level and stable - not to exceed a maximum of
2% deviation from evenness is recommended.
• A smooth wall is required when the system is to be wall-mounted. Check the statics of the wall. It
must have sufficient load-bearing capacity (for system weight, see
).
• An atmospherically floor drain with a outer diameter of 63 mm or 2.48 inch (DN 50 tube) shall be
provided. When no floor drain is available, install a water watcher to protect against water damage
(only available for EU).Otherwise the manufacturer will not accept liability for any possible water
damage.
• Free run off to drain.
Unrestricted gravity flow to drain must be ensured!
• An electrical socket appropriate for the system (see
).
• When installing the ultrapure water system, always ensure that there is adequate space all around
the unit (approx 30 cm / 11.81 inch) to ensure that ease of use or replacement of materials (e.g.
filter change, connection) is possible at all times
• Easy access for operation and control of the system.
• Water pre treated such as DI, RO or distillation water connection with 3/4 NPT male thread and
customer supplied shutoff valve.
The operator is obliged to ensure, that the installation of the water purification unit
and its operation are carried out in compliance with all national and international
guidelines, applicable and valid for the place of installation. If necessary, measures to
protect the drinking water have to be taken by installing appropriate components.
Содержание Barnstead GenPure
Страница 10: ...Contents 4 GenPure Thermo Scientific...
Страница 16: ...2 Safety precautions 10 GenPure Thermo Scientific...
Страница 22: ...4 Intended Use 16 GenPure Thermo Scientific...
Страница 28: ...6 The Installation Area 22 GenPure Thermo Scientific...
Страница 38: ...7 Installation Installation examples 32 GenPure Thermo Scientific...
Страница 40: ...8 Flow charts 34 GenPure Thermo Scientific...
Страница 42: ...9 How the system functions 36 GenPure Thermo Scientific...
Страница 48: ...11 Operating elements Flow chart of menu system control 42 GenPure Thermo Scientific...
Страница 66: ...12 System control GenPure Printer output 60 GenPure Thermo Scientific...
Страница 84: ...13 Maintenance Change and autoclave the Final filter 78 GenPure Thermo Scientific...
Страница 86: ...14 Waste disposal 78 GenPure Thermo Scientific...
Страница 90: ...15 Trouble shooting 82 GenPure Thermo Scientific...
Страница 91: ...Thermo Scientific GenPure 83 16 Replacement parts 10 13 11 4 3 6 5 7 2 9 14 8 1 12 16...
Страница 94: ...17 Consumable materials 86 GenPure Thermo Scientific...
Страница 96: ...18 Accessories 88 GenPure Thermo Scientific...
Страница 97: ...Thermo Scientific GenPure 89 19 Terminal assignments...
Страница 98: ...19 Terminal assignments 90 GenPure Thermo Scientific...
Страница 100: ...20 Maintenance records 92 GenPure Thermo Scientific...
Страница 105: ...Thermo Scientific GenPure 97 I ndex...