13
Maintenance
Maintenance intervals
62
GenPure
Thermo Scientific
Regular servicing of your system ensures that the quality of water is maintained. We recommend a
service contract with a factory authorized service company to ensure that the system is properly
maintained. You then have the certainty of a high operational, safe, and reliable water purification
system.
To ensure error-free operation, your system must be checked, serviced and cared for at regular time
intervals in accordance with these operating instructions. For this reason, the operating instructions
must be readily available to operating and maintenance staff at all times, and be carefully followed.
Calibration of the conductivity is only to be carried out and recorded by a factory-authorized service
technician.
Cleaning and disinfection should be performed at least once yearly, or when the ultrapure cartridge is
replaced, or when bacteria is present in the product water.
Control and maintenance work on electrical systems are only to be carried out by an
appropriately trained, skilled electrician.
Unplug the system from the power outlet for all Maintenance work on the system.
Maintenance intervals
Consumable materials are to be replaced according to the directions below. The intervals were
determined for the average user and are completely dependent on the actual feed water quality and
volume of water used daily.
Material
Flow chart no.
Catalog no.
Interval
Other problems
Ultrapure
cartridge
F1
09.2005
Up to 12
Months
Or when the ultrapure water limiting
value is exceeded, whichever is shorter.
Longer usage can result in bacterial growth
on the resin.
Sterile 0.2 micron
filter
F2
09.1003
Up to 12
Months
Or flow rate is noticeably slower.
Ultrafiltration
membrane (only
applicable for
systems with a UF
filter)
F3
50133980
Up to 24
Months
is markedly slower.
UV-lamp (only
applicable for
systems with a UV
lamp)
UV1
09.2002
Up to24
Months
Or unless system indicates the lamp needs
to be replaced.
*Please keep in mind that the life of your consumable is directly dependent on the quality of the feed
water and the amount of water used daily.
Or if there is endotoxin breakthrough in
product water or when the water flow rate
Содержание Barnstead GenPure
Страница 10: ...Contents 4 GenPure Thermo Scientific...
Страница 16: ...2 Safety precautions 10 GenPure Thermo Scientific...
Страница 22: ...4 Intended Use 16 GenPure Thermo Scientific...
Страница 28: ...6 The Installation Area 22 GenPure Thermo Scientific...
Страница 38: ...7 Installation Installation examples 32 GenPure Thermo Scientific...
Страница 40: ...8 Flow charts 34 GenPure Thermo Scientific...
Страница 42: ...9 How the system functions 36 GenPure Thermo Scientific...
Страница 48: ...11 Operating elements Flow chart of menu system control 42 GenPure Thermo Scientific...
Страница 66: ...12 System control GenPure Printer output 60 GenPure Thermo Scientific...
Страница 84: ...13 Maintenance Change and autoclave the Final filter 78 GenPure Thermo Scientific...
Страница 86: ...14 Waste disposal 78 GenPure Thermo Scientific...
Страница 90: ...15 Trouble shooting 82 GenPure Thermo Scientific...
Страница 91: ...Thermo Scientific GenPure 83 16 Replacement parts 10 13 11 4 3 6 5 7 2 9 14 8 1 12 16...
Страница 94: ...17 Consumable materials 86 GenPure Thermo Scientific...
Страница 96: ...18 Accessories 88 GenPure Thermo Scientific...
Страница 97: ...Thermo Scientific GenPure 89 19 Terminal assignments...
Страница 98: ...19 Terminal assignments 90 GenPure Thermo Scientific...
Страница 100: ...20 Maintenance records 92 GenPure Thermo Scientific...
Страница 105: ...Thermo Scientific GenPure 97 I ndex...