Document 1-0479-013 Rev B
18
Data word length and parity selections available:
Menu
Setting
Data Bits
Parity
Stop Bits
0
8
None
1
1
7
Even
1
2
7
Odd
1
3
8
None
2
4
8
Even
1
5
8
Odd
1
Table 5: Data Word Length and Parity Selections
(3) ModBus RTU
Support for ModBus RTU protocol is available. The default setting for ModBus RTU is
0
(not enabled).
To enable the ModBus RTU mode:
1. In the
Set Baud Rate
menu, enter
(B)
and then select
1
.
.
2. In the
Set Multipoint Parameters
menu, set
Master Packet Repeat
to
3
.
Note: When using the transceivers in ModBus RTU mode the Master Packet Repeat must
be set to 3 regardless of whether the transceivers are being used in Point-to-Point or
Multipoint mode.