Thermo Scientific Barnstead GenPure with xCAD
Ultrapure water system
34
10. Putting the system into operation
Allow the system to warm up to, or to cool down to, room temperature
before starting it up for the first time.
Check that all hose connections have been made as specified in the
"Setting up the system" section.
Switch the system on by pressing this button. After a compulsory rinse the sys-
tem switches to the last operating mode.
To vent air from the system, switch the system 3 times successively to
"Rinsing" in the menu and collect and discard approx. 5 litres of water
each time. The ultra pure water limiting value can be gone below during
this procedure.
Use this “NONSTOP“-key to switch the system to the “Nonstop“ operating
mode.
You can set the system back to “Interval“ operation when the system has
brought the ultra pure water to the quality you require.