9.1 Installation area
Take the following criteria into consideration when selecting the installation area:
Feedwater pressure at least 0,1 bar, at most 6 bar
Lowest tempe2°C
Level standing surface
When the system is to be wall-mounted: A smooth wall is required. Check the statics
of the wall, it must be sufficiently loadable (for weight, see Technical specifications)
Floor drain, DN 50
Free gravity fall to drain.
Should no such floor drain be available, then, for safety reasons, a water watcher (ar-
ticle no. 16.0129) must be installed!
Free gravity fall to drain must be ensured !
Safety socket, 100 - 240V, 50/60Hz
Sufficient working space around the system (filter replacement, etc.)
It must be easy to operate and monitor the system
R 3/4" Tap water connector