7) Lead the 25 pin plugs/extension cables and 4 mm and 6 mm o.d. hoses through the wall
and connect the 25 pin plug (1) from the GenPure with the socket (1) on the xCAD and the
25 pin plug (2) from the GenPure with the socket (2) on the xCAD.
Caution:
Do not interchanging the connections of the 25 pin connector, otherwise no ab-
straction of water possible!
8) Fix the 6 mm und 4 mm o.d. hoses in the hose connectors (5) and 6) of the xCAD.
9) If you intend to use the sterile filter (10) that is standardly supplied, screw it in the R 1/4"
female thread outlet of the dispensing valve (9) of the xCAD.
10) Connect the mains plug (8) of the voltage supply to an earthed 100-240V/50/60Hz socket.
11) Open the feedwater supply.
12) Check all connections for leaks.
9.4 Setting up the system / wall version xCAD
Proceed as follows to set your GenPure with xCAD (wall version) ultra pure water
system up ready for use:
View from below/front
112
Wandhalter
wall mount
Schrauben 4x12 mm
screws 4x12 mm
10
10
132
61
84
24
4xØ6
,3
xCAD mit montiertem Wandhalter
xCAD with mounted wall mount
1) Release the four screws (4x12 mm) marked above from the xCAD underside and