12.2 OEM-Menu:
Basic settings and limiting values can be changed in this menu.
To make changes in the OEM menu, the system control must previously be unlocked (see
section “Menu” Unlocking the system).
Calling the OEM menu:
Simultaneous pressing of the Interval-key and the Nonstop-key calls the OEM menu. Fol-
lowing this, the prompt
"OEM menu Press Enter"
appears. When this is confirmed with the
Enter-key, the first menu prompt can be worked on. To simplify changing settings, press
the UV-key to select the number in a numerical value which you want to change. Now use
the arrow keys to enter the wanted number from 0 to 9 at that selected position.
A press on the menu-key takes you to the next menu prompt.
The OEM menu call display shows:
OEM Menu
12.2.1 Setting the maximum operating temperature:
The maximum temperature which the system is to be allowed to reach can be set in this
menu. When this temperature is exceeded, the "
max. Temperature"
fault message is trig-
gered.
It is shown in the 4th line of the display.
Basic setting:
50 °C
Setting range:
1 - 50 °C
The display shows:
OEM Menu
Max. temp.
35 °C