Operation
Calibration
Thermo Scientific
48iQTL Instruction Manual
3-13
The Calibrate Span Coefficient menu contains the following information:
●
Edit Span Concentration:
User enters the span concentration when in
single range mode.
●
Edit High Range Span Concentration:
User enters the high range span
concentration when in dual or auto range mode.
●
Edit Low Range Span Concentration:
User enters the low range span
concentration when in dual or auto range mode.
●
Current High Range Concentration:
Read only. Current high range
concentration reading when in dual or auto range mode.
●
Current Low Range Concentration:
Read only. Current low range
concentration reading when in dual or auto range mode.
●
Current Span Coefficient:
Read only. Displays what the current user-set
span coefficient is when in single range mode.
●
Current High Range Span Coefficient:
Displays what the current user-set
high range span coefficient is when in dual or auto range mode.
●
Current Low Range Span Coefficient:
Displays what the current user-set
low range span coefficient is when in dual or auto range mode.
●
Calculated Span Coefficient:
Read only. After the “Edit Span
Concentration” value is entered, the new calculated span coefficient is
displayed when in single range mode.
●
Calculated High Range Span Coefficient:
Read only. After the “Edit
High Range Span Concentration” value is entered, the new calculated
high range span coefficient is displayed when in dual or auto range
mode.
●
Calculated Low Range Span Coefficient:
Read only. After the “Edit Low
Range Span Concentration” value is entered, the new calculated low
range span coefficient is displayed when in dual or auto range mode.
●
Calibrate:
When pressed, updates the coefficient and the concentration
should match the span concentration.
Содержание 48iQTL
Страница 1: ...48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014 00 27Sep2018 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 48iQTL Instruction Manual Thermo Scientific ...
Страница 142: ......
Страница 153: ...Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4 11 ...
Страница 162: ...Maintenance Pump Rebuilding 5 4 48iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 180: ...Servicing Fan Replacement 7 8 48iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 208: ...Servicing DMC Optical Bench 7 36 48iQTL Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 221: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 49 Figure 7 53 Heater Assembly Removal pt 4 ...
Страница 223: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 51 Figure 7 55 Heater Assembly Install pt 2 ...
Страница 230: ......
Страница 240: ......
Страница 266: ......
Страница 278: ......
Страница 279: ......