Troubleshooting
Troubleshooting Guide
Thermo Scientific
48iQTL Instruction Manual
6-5
Problem
Possible
Cause
Action
Alarm – Module Thermistor
Open
Defective DMC
board
Replace board as needed.
Alarm – Module Thermistor
Short
Defective DMC
board
Replace board as needed.
Alarm – Bench Temp
Thermistor Open
Cable
connection
Check the cable connection from the
thermistor to the DMC board.
Defective
Thermistor
Replace Thermistor as needed.
Defective
DMC
board
Replace board as needed.
Alarm – Bench Temp
Thermistor Short
Cable
connection
Check the cable connection from the IR
Source to the DMC board.
Defective
IR
Source
Replace IR Source as needed.
Defective
DMC
board
Replace board as needed.
Alarm – IR Source Current
Low
Cable
connection
Check the cable connection from the IR
Source to the DMC board.
Defective
IR
Source
Replace IR Source as needed.
Defective
DMC
board
Replace board as needed.
Alarm – IR Source Current
High
Cable
connection
Check the cable connection from the IR
Source to the DMC board.
Defective
IR
Source
Replace IR Source as needed.
Defective
DMC
board
Replace board as needed.
Alarm – IR Detector Bias
Low
Cable
connection
Check the cable connection from the IR
Source to the DMC board.
Defective
IR
Source
Replace IR Source as needed.
Defective
DMC
board
Replace board as needed.
Alarm – IR Detector Bias
High
Cable
connection
Check the cable connection from the IR
Source to the DMC board.
Defective
IR
Source
Replace IR Source as needed.
Defective
DMC
board
Replace board as needed.
Содержание 48iQTL
Страница 1: ...48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014 00 27Sep2018 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 48iQTL Instruction Manual Thermo Scientific ...
Страница 142: ......
Страница 153: ...Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4 11 ...
Страница 162: ...Maintenance Pump Rebuilding 5 4 48iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 180: ...Servicing Fan Replacement 7 8 48iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 208: ...Servicing DMC Optical Bench 7 36 48iQTL Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 221: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 49 Figure 7 53 Heater Assembly Removal pt 4 ...
Страница 223: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 51 Figure 7 55 Heater Assembly Install pt 2 ...
Страница 230: ......
Страница 240: ......
Страница 266: ......
Страница 278: ......
Страница 279: ......