Servicing
Measurement Side Removal
Thermo Scientific
48iQTL Instruction Manual
7-9
Use the following procedure to remove and replace the measurement side if
necessary.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug the fan cable J12 on optical bench DMC board.
Figure 7–9.
Unplugging the Fan Cable
3.
Unplug DMC cable (
Figure 7–10
).
Measurement
Side Removal