Operation
Settings
Thermo Scientific
48iQTL Instruction Manual
3-67
The Averaging Time screen allows the user to choose dynamic filtering or a
manually selected (static) averaging time. The 48iQTL, by default, uses
dynamic filtering to smooth output data.
Dynamic Filtering allows for data smoothing without compromising
response time. When dynamic filtering is enabled, the instrument software
automatically changes the averaging time giving the user faster response
times when conditions are rapidly changing; smoother and stable readings,
when conditions aren’t changing as rapidly; and as an added bonus, it
better processes spikes to minimize their impact on the data. At the same
time it will preserve the representative nature of the filtered data to the
conditions being sampled.
Averaging Time defines the time period (1 to 300 seconds) during which
CO measurements are taken. The average concentration of the readings are
calculated for that time period. The front panel display and analog outputs
are updated every second if averaging time is greater than 1 second. An
averaging time of 10 seconds, for example, means that the average
concentration of the last 10 seconds will be displayed at each update. An
averaging time of 300 seconds means that the moving average
concentration of the last 300 seconds will be the output at each update.
Therefore, the lower the averaging time the faster the front panel display
and analog outputs respond to concentration changes. Longer averaging
times are typically used to smooth output data.
Note
When Dynamic Filtering is selected, the user selected Averaging
Time button is disabled.
▲
Home Screen>Settings>Measurement Settings>Averaging time (single
range mode)
Averaging Time
Содержание 48iQTL
Страница 1: ...48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014 00 27Sep2018 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 48iQTL Instruction Manual Thermo Scientific ...
Страница 142: ......
Страница 153: ...Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4 11 ...
Страница 162: ...Maintenance Pump Rebuilding 5 4 48iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 180: ...Servicing Fan Replacement 7 8 48iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 208: ...Servicing DMC Optical Bench 7 36 48iQTL Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 221: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 49 Figure 7 53 Heater Assembly Removal pt 4 ...
Страница 223: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 51 Figure 7 55 Heater Assembly Install pt 2 ...
Страница 230: ......
Страница 240: ......
Страница 266: ......
Страница 278: ......
Страница 279: ......