Servicing
DMC Optical Bench
Thermo Scientific
48iQTL Instruction Manual
7-33
Note
Replace the optical switch if needed. You don’t have to remove the
optical switch to replace the motor or filter.
▲
5.
Using #1 Phillips drive, unfasten two #4-40 flat head screws to remove
the optical cover.
6.
Using #1 Phillips drive, unfasten one #2-56 pan head screw to remove
the optical switch.
Figure 7–38.
Motor and Filter Optical Switch Replacement (pt 3)
Содержание 48iQTL
Страница 1: ...48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014 00 27Sep2018 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 48iQTL Instruction Manual Thermo Scientific ...
Страница 142: ......
Страница 153: ...Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4 11 ...
Страница 162: ...Maintenance Pump Rebuilding 5 4 48iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 180: ...Servicing Fan Replacement 7 8 48iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 208: ...Servicing DMC Optical Bench 7 36 48iQTL Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 221: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 49 Figure 7 53 Heater Assembly Removal pt 4 ...
Страница 223: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 51 Figure 7 55 Heater Assembly Install pt 2 ...
Страница 230: ......
Страница 240: ......
Страница 266: ......
Страница 278: ......
Страница 279: ......