Servicing
Filter Replacement
7-6
48iQTL Instruction Manual
Thermo Scientific
Use the following procedure to replace the filter.
1.
Turn instrument OFF and unplug the power cord.
2.
Starting with top right corner, pull out to remove fan cover.
Figure 7–5.
Start with Top Right Corner of Fan Cover
3.
Replace filter and snap fan cover back in place.
Figure 7–6.
Removing the Fan Cover
Filter
Replacement