Operation
Settings
3-78
48iQTL Instruction Manual
Thermo Scientific
The Compensation screen provides compensation for any changes to the
instrument’s output signal due to pressure.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Compensation
The Compensation screen contains the following information:
●
Pressure Compensation:
Toggles pressure compensation enabled or
disabled and provides compensation for any changes to the
instrument's output signal due to bench pressure variations. The effects
of bench pressure changes on the analyzer's subsystems and output have
been empirically determined. This empirical data is used to compensate
for any change in chamber pressure.
●
Standard Temperature for Gas Units:
Allows the user to enter a value
between 0 and 30
°
C to correct the
g/m
3
, mg/m
3
, and g/m
3
concentration calculations.
Compensation
Содержание 48iQTL
Страница 1: ...48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014 00 27Sep2018 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 48iQTL Instruction Manual Thermo Scientific ...
Страница 142: ......
Страница 153: ...Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4 11 ...
Страница 162: ...Maintenance Pump Rebuilding 5 4 48iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 180: ...Servicing Fan Replacement 7 8 48iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 208: ...Servicing DMC Optical Bench 7 36 48iQTL Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 221: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 49 Figure 7 53 Heater Assembly Removal pt 4 ...
Страница 223: ...Servicing Heater Assembly Thermo Scientific 48iQTL Instruction Manual 7 51 Figure 7 55 Heater Assembly Install pt 2 ...
Страница 230: ......
Страница 240: ......
Страница 266: ......
Страница 278: ......
Страница 279: ......