TLC-
2000
THERAPEUTIC MEDICAL LASER
Document: 600-0013
36
Recommended separation distances between portable and mobile RF communications equipment
and the TLC-2000
The TLC-2000 is intended for use in the electromagnetic environment in which radiated RF disturbances are con- trolled.
The customer or the user of the TLC-2000 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the TLC- 2000 as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
80 M kHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.2
1.2
2.3
10
0.37
0.37
0.74
100
11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance in meters(m) can be
estimated using the equation applicable to the frequency of the transmitter, where is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.