![THERALASE TLC-2000 Series Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/theralase/tlc-2000-series/tlc-2000-series_operation-manual_1101762012.webp)
TLC-
2000
THERAPEUTIC MEDICAL LASER
Document: 600-0013
1
0
Power supply TLC-
2003
plugs in to
the Power Pack - TLC-
2002
Power output: Voltage: +24 VDC
Current: 4.0 A
Plugs in to main AC line via power
cord (TLC-155)
C
aution:
•
Use only the TLC-2003 Power Supply (+24 VDC, 4.2 A) to power the
TLC-2002 Power Pack. This is the only Power Supply approved to be used
with the TLC- 2000 laser system.
•
Outside of North America, the proper international cord set must be
specified and purchased from Theralase™ Inc. or purchased separately for
the country of preference.
The Power Supply (TLC-
2003
)
The TLC-
2003
is a universal power supply – the input voltage range is 100VAC to 240VAC, 47 to 63 Hz, meaning that it
can be safely used anywhere in the world. The TLC-
2003
power supply is able to power the TLC-
2002
Power Pack.
Figure 6. TLC-
2003
Power Supply and Associated Line Cord.
With the proper cord set, the TLC-
2003
can be plugged in anywhere in the world. The cord set (TLC-
155
) supplied with
the Theralase™ TLC-
2003
power supply is a hospital grade 115 to 120 VAC, 50 to 60 Hz approved cord set. This is built to
a North American standard and was chosen to work in Canada and the United States.
Some of the key features of the TLC-2003 Power Supply include:
100-240 VAC Universal Input
Meets Safety Agency Requirements
In Case IEC320
Complies with EMC/EMI Regulations
Single Output up to 101 Watts
CE Compliant
Outputs Regulated with Low Ripple Impact Resistant Polycarbonate Enclosure