5) MAINTENANCE AND SERVICE

This section deals with maintenance and serviceability of the Theralase™ system by the end user. If further detailed service of your 

laser system is required, contact Theralase™ Inc directly for assistance. Detailed contact information is supplied at the end of this 

section. 

Inspection and Cleaning 

PRE AND POST TREATMENT 

Wipe the surface of the probe with a 70% solution of isopropyl alcohol. A 70% solution of isopropyl alcohol is a universally 

accepted disinfectant in accordance with CDC (Center for Device Control and Prevention) Guideline for Disinfection and 

Sterilization in Healthcare Facilities, 2008. William A. Rutala, Ph.D., M.P.H.

1,2

, David J. Weber, M. D., M. P. H. 

1,2

, and the 

Healthcare Infection Control Practices Advisory Committee (HICPAC)

3

.

 

WEEKLY 

Follow the stop procedure in the basic operation section of this manual. Remove the key from the Power Pack. Disconnect 

the Laser Probe(s) and power input to the Power Pack. Carefully clean each of the TLC-2001 Laser Probes with isopropyl 

alcohol swabs being careful not to immerse the Laser Probe heads. Remove the lens cap and immerse a Q-tip into isopropyl 

alcohol and rotate in the laser aperture carefully to avoid damaging the laser diodes. 

MONTHLY 

Follow the stop procedure in the basic operation section of this manual. Remove the key from the Power Pack. Disconnect 

the Laser Probe(s) and power input to the Power Pack. Inspect the laser diodes for cracks or chips in the glass. A visible chip 

in the glass, a crack in the glass or a missing glass mandates replacement of the laser diode. Contact Theralase™ 

immediately for replacement of the defective / damaged laser diode. 

PREVENTIVE MAINTENANCE SCHEDULE 

In addition to performing the monthly inspection for physical problems associated with the laser diodes, it is also important 

to check the efficacy of each of the laser diode outputs. Laser diodes typically wear or drop in average power as the total 

usage hours of the system increases. The real time laser power indicator is used as a reference for when the laser diodes 

should be calibrated / replaced. 

Check the optical output of each laser diode to ensure laser diode optical output at least once per week. 

The TLC-2000 therapeutic laser system may only be calibrated/repaired by Theralase™ Inc. or trained personnel equipped 

with the proper service equipment. There are no user serviceable parts in the Laser Probe. The Laser Verification software 

will indicate if a laser diode service must be performed. 

Transport 

Remove the key from the main power switch and store in a safe location. Disassemble the laser system, carefully packing all 

components and taking special care with the laser probe assembly. In packing and unpacking the apparatus, be careful not to 

twist or kink the probe cable(s) too sharply to prevent damage to the probe cable assembly. Ensure proper storage of the 

Laser Probes to avoid laser diode damage. If you damage your laser diodes, please notify Theralase™ Inc. immediately for 

service / replacement. (Note: The laser system should be kept within 0 to 40°C, 500 hPa to 1068 hPa and 0 to 80% relative 

humidity during transport). 

Содержание TLC-2000 Series

Страница 1: ......

Страница 2: ...ications in the use of Laser Therapy 18 Cautions in the use of Laser Therapy 18 3 START UP INFORMATION 21 The Power Pack TLC 2002 21 The Laser Probe TLC 2001 22 The Computer Tablet TLC 2004 24 Set Up...

Страница 3: ...Safety and safe use of the product are indicated both visually by the use of labels as well as equipment features These will be detailed below NOTE It is important to be well aware of all these safet...

Страница 4: ...and reliability of the protective earth connection Includes an F TYPE APPLIEDPART Laser Radiation Warning Label VISIBLE AND INVISIBLE LASER RADIATION AVOID EXPOSURE TO BEAM 905nm LASER PULSE DURATION...

Страница 5: ...lock Connector is a connector that may be removed for disconnecting the main power 4 Laser Probe The Laser Probes are the applied parts of the system The Probe cannot operate the laserdiodes unless a...

Страница 6: ...and can be ordered Part TLC 901 directly from Theralase Inc The laser eyewear is for diffuse viewing of laser light only The Theralase TLC 2000 laser output is athermic Laser diode package cases may...

Страница 7: ...from water intrusion and should never be used wet or in the accompaniment of water If the laser system becomes wet or immersed in water thoroughly dry the laser system and allow 48 to 72 hours in a w...

Страница 8: ...symptoms The TLC 2000 therapeutic laser system is not promoted as a cure for any patient ailments The information contained in this manual is for reference only and does not preclude the practitioner...

Страница 9: ...ower Pack only supplies power to the Laser Probe s The Laser Probes are controlled by Theralase software resident on the Tablet Under software control the Laser Probes operate in a patient treatment m...

Страница 10: ...ariable Total Number of Visible Lasers 4 Total Visible Laser Power 400 mW COMPLIANCES North America CAN CSA C22 2 NO 60601 1 08 CAN CSA C22 2 NO 606001 2 10 14 CAN CSA C22 2 NO 606001 2 22 08 ANSI AAM...

Страница 11: ...e unit has an emergency STOP button red button with white arrows The front of the unit has a key switch for switching the unit off and prohibiting the unit from unauthorized use an interlock switch an...

Страница 12: ...age range is 100VAC to 240VAC 47 to 63 Hz meaning that it can be safely used anywhere in the world The TLC 2003 power supply is able to power the TLC 2002 Power Pack Figure 6 TLC 2003 Power Supply and...

Страница 13: ...en the time setting is displayed will the start pushbutton have any effect The black pushbutton may be operated once to turn on the laser diodes for the time setting displayed on the Laser Probe The o...

Страница 14: ...Assembly illustrating the key Note that the key is located at the 6 o clock position The cable plug assembly must be properly aligned with the receptacle assembly and then inserted until the latches e...

Страница 15: ...ng laser therapy treatment Figure 11 Theralase supplied Laser SafetyEyewear Optical Density O D is the scientific term to describe a material s ability to block light of a certain wavelength It isa lo...

Страница 16: ...0 mW average power and 905 nm 200 mW average power laser diodes Therefore using the Theralase approved laser glasses may result in the following levels of light being transmitted to the eye assuming o...

Страница 17: ...allow 2 weeks as a wash out period prior to laser treatment If a wash out period is not possible reduce the laser power to 1 2 of the recommended power indicated and increase with each treatment as th...

Страница 18: ...and pull the Remote Interlock Connector straight out Main power is now disabled and no action will restore it until the Remote Interlock Connector is replaced To replace insert straight into the match...

Страница 19: ...e There are different strap sizes provided with the Laser Probes to facilitate holding the probe on different aspects of theknee Note Pressing the black pushbutton on the side of the TLC 2001 Laser Pr...

Страница 20: ...output window is an optically transparent acrylic plastic mounted to a ring The window is removable by turning the ring left until it stops and lifting the ring straight off the probe The window is co...

Страница 21: ...d that the system be configured for new passwords and user names Refer to the section Setting Username and Passwords in the Administrator Menu section for thisprocedure The Computer Tablet is shipped...

Страница 22: ...LC 2004 2 Tablet power supply and cable 3 Power Supply TLC 2003 4 Medical AC power cable 5 Power Pack TLC 2002 6 Laser Probe s 1 to 4 TLC 2001 7 Probe cable s 1 to 4 8 Safety Eyewear Figure 16 TLC 200...

Страница 23: ...source Thepower supply is universal and will need a plug adaptor for 240 VAC 6 Open the Computer Tablet and turn it on Then log into the operating system with the supplied username and password 7 Sele...

Страница 24: ...Black Start pushbutton once on the Laser Probe This will determine an optimal protocol for the particular patient s physical profile 2 Having already completed a measurement select the Treatment Mode...

Страница 25: ...obe or the Start button directly on the software itself Stop Procedure The Laser Probe may be turned off during operation by any of the following means 1 Slide the switch on the Laser Probe to the OFF...

Страница 26: ...to create a custom protocol Enter the power level for 905nm and 660nm treatment duration protocol name and body region to be treated Advanced Treatment Customized Treatment Options Basic Treatment Pr...

Страница 27: ...s used to configure Probes to the TLC 2000 software via BlueTooth System Operation in Detail Common Operations Turning the Computer Tablet ON Turn on the tablet by pressing and holding the power on bu...

Страница 28: ...sktop with the Theralase TLC 2000 application icon Launching the TLC 2000software Tolaunch the Theralase application double touch the icon and the application Sign In window willappear Enter the usern...

Страница 29: ...he To Diagnoses button to proceed to the Patient Treatment Diagnoseswindow For existing diagnoses select the diagnosis to be treated and select the Start New Treatment button For a new diagnosis selec...

Страница 30: ...he Treatment Mode The Laser Probe will Beep and go into the ARMED state Place the Laser Probe on the area of the patient to be treated Start the treatment by pressing the Start button on the software...

Страница 31: ...ient drop down menu Hit the save button You will see your new custom protocol appear on the right hand side of the screen To create a new custom protocol simply hit the plus button again To delete a c...

Страница 32: ...for one or two users Save the data To remove a user highlight the name from the list and touch the trash canicon To Adjust Probe Volume Select the Probes Tab You can adjust the volume of the probe s...

Страница 33: ...acks or chips in the glass Avisible chip in the glass a crack in the glass or a missing glass mandates replacement of the laser diode Contact Theralase immediately for replacement of the defective dam...

Страница 34: ...of delivery This warranty includes all parts and labour used in the repair of the TLC 2000 laser system for the term of five 5 years This warranty is considered null and void if the laser system has...

Страница 35: ...s used in such anenvironment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The TLC 2000 uses RF energy only as a bi product of its internal function Ther...

Страница 36: ...rt interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 95 dip in for 0 5 cycles 40 60 dip in for 5 cycles 70 30 dip in for 25 cycles 5 95 dip in for 5 s 5 95 dip in for 0...

Страница 37: ...d RF transmitters as determined by an electromagnetic sitesurveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinityof equipment marked with the f...

Страница 38: ...put power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 M kHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 2 1 2 2 3 10...

Отзывы: