TLC-
2000
THERAPEUTIC MEDICAL LASER
Document: 600-0013
18
Caution:
NEVER look directly into this area with the lasers on.
4. Start/Stop Button: This button is used to start/stop the Laser Probe laser light emission. Press this button once to
commence emission of laser light. Pressing the button once, while the probe is operating, will suspend the
countdown and turn off laser emission. The operation and countdown can be resumed by pressing the button in the
correct start sequence again.
5. Laser Diode Aperture: This area is where the laser energy is emitted and is the treatment contact area. There are
five (5) 905 nm invisible, Near Infrared 200 mW maximum average power laser diodes and four (4) 660 nm, visible
100 mW maximum average power, red laser diodes.
6. Laser Output Window: The laser output window is an optically transparent acrylic plastic mounted to a ring. The
window is removable by turning the ring left until it stops and lifting the ring straight off the probe. The window is
coated to be scratch resistant. It keeps the lasers clean and can be wiped with isopropyl alcohol. Windows that
become scratched should be replaced. Replacement windows can be ordered directly from Theralase™.
7. Laser Probe Cover: The Laser Probe cover attaches to the probe by magnets. The cover protects the window
during storage. It is recommended that the cover be installed when the Laser Probe is not in use.
8. Probe Body: The Laser Probe body is molded from Bayblend FR2000, a medical grade, fire retardant PVC
material. This material is impervious to most cleaning and sterilizing solutions, and is cleaned with isopropyl alcohol.
9. Strap Handles: Straps to hold the Laser Probe in place on the treatment area attach to these handles on the sides
of the Laser Probe.
1
3
4
6
5
7
2
8
9