![THERALASE TLC-2000 Series Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/theralase/tlc-2000-series/tlc-2000-series_operation-manual_1101762035.webp)
TLC-
2000
THERAPEUTIC MEDICAL LASER
Document: 600-0013
3
3
6) ELECTRO MAGNETIC COMPATIBILITY INFORMATION
In accordance with clause # 5.2.x of EN 60601-1-2.
Interconnecting Cables
Cable NO.
Type
Length
Connected From
Connected To
Special Accessories (ferrites, etc.)
1
Power Cord
2.37M
Power Supply
Power Outlet
None
Warning: The use of Power Cord other than the one provided by Theralase™ Inc. for the TLC-2000 may result increased
EMISSIONS or decreased IMMUNITY of the TLC-2000 system.
Warning: The TLC-2000 should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is
necessary, the TLC-2000 should be observed to verify normal operation in the configuration in which it will be used.
Guidance and THERALASE™ INC’s declaration- ELECTROMAGNETIC EMISSIONS
The TLC-2000 is intended for use in the electromagnetic environment specified below. The customer or the user of the
TLC-2000 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The TLC-2000 uses RF energy only as a bi-product of its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronics
equipment.
RF emissions CISPR 11
Class B
The TLC-2000 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply network that
supplies buildings used for domestic purposes.
Guidance and THERALASE™ INC’s declaration- electromagnetic IMMUNITY
The TLC-2000 is intended for use in the electromagnetic environment specified below. The customer or the user of the
TLC-2000 should assure that it is used in such an environment.
IMMUNITY test
IEC60601 Test level
Compliance level
Electromagnetic environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6KV contact
±
8KV air
±
6KV contact
Criteria A
Level 3
±
8KV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%
Содержание TLC-2000 Series
Страница 1: ......