THERAPEUTIC MEDICAL LASER
Document: 600-0013
1
1
Caution: The Laser Probe Cover should be placed over the lens area when not in
Laser Probes (TLC-
2001
)
The TLC-
2002
Power Pack is designed to operate with up to four (4) of the TLC-
2001
Laser Probes. The TLC-
2001
Laser Probe has a POWER ON slide switch and a BLACK pushbutton START switch. The Lasers cannot be activated
until the Laser Probe receives a command transmission from the Tablet. The Probes do not retain any clinical protocols in
memory when disconnected from power.
When first turned on, the Laser Probe will be in idle mode waiting for data and a command transmission from the Tablet.
The Laser Probe will display the treatment time on the liquid crystal display (“LCD”) when the information is received from
the Tablet. Only when the time setting is displayed will the start pushbutton have any effect. The black pushbutton may be
operated once to turn on the laser diodes for the time setting displayed on the Laser Probe. The operating status of the
Laser Probe will be displayed on the LCD. The display will change the Laser status to RED while the laser diodes are
emitting and the time duration will count down to zero. The fans will turn on during the time the lasers are on and will stop
when the time reaches zero and the laser diodes are turned off.
The Laser Probe has to be used in such a manner to encourage free air flow around it. This would discourage placing the
Laser Probe under gowns, toweling or clothing if it is not absolutely necessary.
Figure 7. Top and Bottom view of the Laser Probe (TLC-2001).
The output of the TLC-2001 Laser Probe consists of:
•
five (5) 905 nm, invisible, Near Infrared 200 mW maximum average power laser diodes
•
four (4) 660 nm, visible, red 100 mW maximum average power laser diodes