11-1-8. Local bus connection
The local bus can control multiple PS-A power supply units from a single communication port on a PC.
A PC or sequencer must be connected to the PS-A power supply unit (for which "1" is the system address) that is set
s the master unit. Up to 31 slave units can be connected in a cascade via dedicated modular cables.
The use of RS-485 signals allows the distance between PS-A units to be increased. However, performance may be
delayed because the master unit converts commands.
(1) When IF-70GU is used:
a
ave units are designated by using the ADRS command to specify the system address numb
Sl
ers.
The IF-70GU interface setting of the slave units can be either USB or GP-IB.
IF-71RS for which RS-232C has been specifie
An
Onl
e on. The other terminators are off.
(2) W
d can be used as the interface for slave units.
y the terminators at both ends of the local bus should be on. The rest should be off.
he above chart, the termin
In t
ators for system addresses "1" and "4" ar
hen IF-71RS is used:
PC
IF-71RS
PS-A power supply
(Slave unit)
System address
: 2
Terminating resistor
IF-71RS
PS-A power supply
(Master unit)
System address
RS-232C mode
: 1
Terminating resistor: ON
COM1
IF-71RS
PS-A power supply
(Slave unit)
System address
: 3
Terminating resistor
IF-71RS
PS-A power supply
(Slave unit)
System address
: 4
Terminating resistor: ON
Local bus
Local bus
Local bus
RS-232C mode
RS-232C mode
RS-232C mode
Slave units are designated by using the ADRS command to specify the system address numbers.
Slave units in which IF-70GU boards have been installed can be used.
Only the terminators at both ends of the local bus should be on. The rest should be off.
In the above chart, the terminators for system addresses "1" and "4" are on. The other terminators are off.
11-1-9. When used as a PSR or PSR-M supply unit
When the appropriate settings are specified for the IF-71RS, the PS-A series units are compatible with and can be
used in place of our PSR series and PSR-M series.
For a details about the commands, see the respective PSR series and PSR-M series instruction manuals.
The IF-71RS settings are as follows:
Operation mode
CN5
CN6
CN7
CN8
Connector to be used
PSR mode
Closed
Open
Open
Closed
※
2
PSR
BUS
PSR-M mode
Closed
※
1
Closed Closed Closed
※
2
PSR-M
BUS
※
1
Only the connectors at both ends of local bus should be closed.
The terminators should be on.
※
2
Open the connector if a ground loop causes a malfunction to occur.
81