8-4. Confirming the Setting Items for Steps
You can confirm the step setting items.
Operation procedure
1. Press the CHECK key, which is then lit in green.
The voltage and current indicators display the
step setting items.
2. While the STOP key or PAUSE key is lit in amber
(while the sequence program is stopped or
paused), turn the rotary encoder clockwise 1
click to switch between the different step setting
items. The setting items are displayed in the
voltage and current indicators.
See the table below for details about the step
setting items.
Voltage indicator
Current Indicator
Setting item
Remarks
Set voltage
Set current
Voltage/current
on
oFF
Out
Hi-r
OUTPUT
When the output HI-R function is on, the output is
set to off.
Set OVP
oVP
OVP value
Set UVP
uVP
UVP value
Set OCP
oCP
OCP value
H n
Execution time
(hr, min)
Sec. m
sec.
Execution time
(sec, m sec)
The execution time can be set to the m sec.
However, the set time is rounded up to 10m sec.
when applying setting items.
If the execution time is 0 hr 0 min 0 sec 0m sec,
the step is skipped and the setting items are not
executed.
PAuS oFF PAUSE
If the pause setting is on, the sequence program
will pause when the step is finished.
cc-P on/oFF
CC priority output
setting
While a sequence program is being executed automatically (while the START LED is lit in amber), only the set voltage
and current can be confirmed. The other step setting items cannot be confirmed.