11
INSTALLATION
5.3
Installation Procedure
Preparing the sealer for use
1.
Fit correct mains input plug to the Battery charger.
2.
The changeable mains connector is available for European continent (standard), UK,
USA/Canada and Australia/New Zealand and as IEC standard 320 C8.
3.
Connect the charger (grey on connector) to the combined Hand unit/Battery charger
cable connector.
4.
Plug the charger into the mains. The charger can be connected to mains voltages
between 100 and 240 V AC.
5.
It takes about three hours to charge the Battery unit completely. When the charging
level indication on the unit indicates 100% and the recharge indication flashes, the
battery is fully charged.
6.
Charge the battery to 100% before the first use.
7.
Connect the Hand unit cable (grey at end of the cable) into the Battery unit connector.
Make sure the connector snaps in secured position.
8.
Perform a test seal on an empty or water filled tube to ensure proper operation.
CAUTION
Only use the enclosed Battery charger for charging the Battery unit.
NOTE
Ensure that cable to handle is connected before use.