26
EMC TABLES
Table 6:
Recommended separation distance between portable and mobile
RF communications equipment and the T5460 T-SEAL MOBILE
The T5460 T-SEAL MOBILE is intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the T5460 T-SEAL MOBILE can help prevent electro-magnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the T5460 T-SEAL MOBILE as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.17
√
P
80 MHz to 800 MHz
d = 1.17
√
P
800 MHz to 2.5 GHz
d = 2.33
√
P
0.01 0.12
0.12
0.23
0.1 0.37
0.37
0.74
1 1.2
1.2
2.3
10 3.7
3.7
7.4
100 11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer
NOTES
At 80MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.