16
OPERATING INSTRUCTIONS
CAUTIONS
In case of sealer malfunction (intermittent operation, poor seal quality,
the sealing time seems too long or too short) contact your local
Terumo BCT representative for assistance.
Periodically check whether the shape of the seal is symmetrical and
there is a clear line in the middle of the seal (see picture below).
Good sealing pattern Bad
sealing
pattern
NOTES
The tube must be dry on the outside.
Do not pull the tube and keep the trigger fully pressed down until the
light goes out. If the trigger is released the sealing process stops.
7.3
If the Sealer Doesn´t Start
The Sealer has several safety functions. Before sealing, these functions control and identify
whether it is possible to seal the tube. The table below covers the common probable causes
for problems and suggests some recommended actions.
Probable cause
Recommended action
Leaky or wet tube
Dry the tube and try again
Tiny arcs between the electrodes
Dry the electrodes and try again
Wet or dirty electrodes
Clean and dry the electrodes
Low/no capacity in battery
Check/charge battery
Overheated (Hand unit LED red blinking)
Let the Hand unit cool down.
No seals
Check Battery unit and connection.
If the sealer still doesn’t start, see chapter 8 for further information
7.4
LED indications
Battery unit
•
“Colored LEDs bar indicating battery status.”
•
“TEST”: button to check the battery. When pressed, the battery charge level indicator
lights up. Refer to section 6.3.
Hand unit
•
Indicator lights orange when RF-energy is being applied to the fixed electrode and
turns off when sealing is completed.
•
Indicator flashes red at overheating, sealing function is blocked.
•
Indicator lights red one second in case of short circuit, sealing function is blocked.
•
Indicator lights red in case of battery low level, sealing function is blocked.