background image

 

15

 

 

 

OPERATING INSTRUCTIONS    

 

 

7

 

OPERATING INSTRUCTIONS 

This chapter describes the use of the instrument.  

 

WARNINGS 

T-SEAL MOBILE uses radio frequency (RF) energy to generate heat 
for sealing. Users should be cautious of potential electrical shocks or 
hazards while handling this sealer. Always keep your fingers away 
from the electrodes in the slot. Never place any object other than the 
tube between the electrodes. 

 

 

This machine emits a low level of electromagnetic (non-ionizing) 
radiation while sealing. It should not be used near high frequency 
sensitive electronic equipment. 

See table 1 for guidance. 

 

 

Do not allow the Hand unit to come directly in contact with the donor. 

 

 

Do not perform a seal within 8 cm (3 in) of needle to preclude an RF 
burn at the needle entry point. 

 

CAUTIONS 

Inspect all parts of the instrument for defects before use. Check 
sealing pattern if Hand unit is dropped. 

 

 

Always check the seals when the battery capacity is low (20% or 
less). 

7.1

 

Preparation before Use 

1.

 

Place the instrument on a flat surface near the working place or in the holster. 

2.

 

Check that there is sufficient battery capacity by pushing the “TEST” button. 

3.

 

Connect the Hand unit cable into the combined Hand unit/Battery charger cable 
connector. 

 

NOTE 

Do not try to connect other sealers to the T-SEAL MOBILE Battery 
unit than the Terumo BCT T-SEAL MOBILE Hand unit. 

7.2

 

How to Seal Tubes 

1

 

Put the tube to be sealed down to the bottom of the slot in the Hand unit.  

2

 

Check that the tube is placed between the electrodes in the slot. 

3

 

Press the trigger to bring the two electrodes closer together until the light on top of the 

Hand unit comes on. The sealing time is normally 0.4 to 0.9 seconds; after a 
maximum of 3 seconds the RF is turned off. 

4

 

When the light goes out, sealing is finished. Release the trigger and remove the tubing.  

5

 

The center of the sealed pattern is very thin and pulling both sides will divide the tube 

into two pieces.  

6

 

Check the tube for leakage. 

 

CAUTION 

If you should make two or more seals, they should not be within 1 cm 
(0.4 in) of each other, otherwise the resulting pressure in the tube 
may cause microscopic cracks and holes in the seals. 

Содержание T-SEAL MOBILE

Страница 1: ...LCT 7010en 12 2015...

Страница 2: ...we make no guarantee to this effect and would appreciate any observations regarding the contents of this document We may make improvements and alterations to the instrument and these changes will be...

Страница 3: ...on Procedure 11 6 FUNCTIONAL DESCRIPTION 12 6 1 Description of Sealing 13 6 2 Description of Equipment 13 6 3 Checking the Battery unit 14 7 OPERATING INSTRUCTIONS 15 7 1 Preparation before Use 15 7 2...

Страница 4: ...tubing into the slot at the front of the Hand unit to create a series of seals T SEAL MOBILE is comprised of a Hand unit and Battery unit which is connected to the Hand unit with the integrated cable...

Страница 5: ...arging Battery charger UL recognized component for Canada and U S A Symbol for CATALOGUE NUMBER Symbol for WARNING Follow instructions Symbol for SERIAL NUMBER Symbol for DATE OF MANUFACTURE Symbol fo...

Страница 6: ...be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the equipment or system should be observed to verify normal operation in the configuration in which it will...

Страница 7: ...A tubes connected to blood bags included either in Blood Packs or in Apheresis Disposable Sets The sealing can be performed when the donor is still connected to the Blood Pack or the Apheresis Disposa...

Страница 8: ...comprises the parts listed below Description Hand unit Battery unit Battery charger without plug Kit Primary Adaptors Part No 546900000 547900000 544900000 54410120 List of all cables and maximum leng...

Страница 9: ...Accessories Description Holster Extension cord 1 2 m 5 pol REF 8METSM004 8METSM005 4 4 User Interchangeable Parts Description Hand unit Battery unit Battery charger without plug Kit Primary Adaptors R...

Страница 10: ...e above list is subject to change refer to the packing list for accurate description of contents If any parts are missing or if the parts are damaged report it immediately 4 Please keep all shipping a...

Страница 11: ...mains voltages between 100 and 240 V AC 5 It takes about three hours to charge the Battery unit completely When the charging level indication on the unit indicates 100 and the recharge indication fla...

Страница 12: ...ectors and the indications are placed on the unit and their functions Front view of Battery unit Charge level indicator Recharge indicator Battery test button Hand unit Battery charger cable connector...

Страница 13: ...it detects controls and adjusts the necessary sealing activity to give the best sealing quality for the type of tubes that are being used See section 7 4 for a description of the different LED indicat...

Страница 14: ...ndicator lights up When the recharge indicator starts flashing the battery is fully charged and the charge current is automatically turned off The charge can be interrupted at any time without damage...

Страница 15: ...e 1 Place the instrument on a flat surface near the working place or in the holster 2 Check that there is sufficient battery capacity by pushing the TEST button 3 Connect the Hand unit cable into the...

Страница 16: ...ests some recommended actions Probable cause Recommended action Leaky or wet tube Dry the tube and try again Tiny arcs between the electrodes Dry the electrodes and try again Wet or dirty electrodes C...

Страница 17: ...oo low voltage Charge battery for 3 hours Fuse blown Contact your local Terumo BCT representative Other cause Contact your local Terumo BCT representative Charge Level LEDs do not light up during char...

Страница 18: ...ur local Terumo BCT representative Hard to divide tube after seal Moving electrode doesn t move smoothly in handle Clean and oil electrode shaft see chapter 9 2 Wet electrodes Dry the electrodes and t...

Страница 19: ...tion solution used remember to remove any residue to ensure that surfaces of the equipment are not subject to corrosion or discoloration Discard all materials in contact with blood according to instit...

Страница 20: ...r each cleaning Assemble the parts in reverse order Make sure the trigger and spring lock is in full forward position before electrode insertion Insert the moving electrode ensuring that it is positio...

Страница 21: ...ts including any warranty of merchantability or appropriateness for a particular purpose Terumo BCT s sole obligation and original purchaser s exclusive remedy for breach of warranty shall be limited...

Страница 22: ...collection and return systems available to you For more information on return recovery or recycling of T SEAL MOBILE please contact your local Terumo BCT Sales Office Technical support Call free numbe...

Страница 23: ...gnetic Environment Guidance RF emission CISPR 11 Group2 The T5460 T SEAL MOBILE must emit electromagnetic energy in order to perform its intended function Nearby electronic equipment may be affected R...

Страница 24: ...uld be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV line s to line s 2kV line s to earth 1kV 2kV Mains power quality should be that of a typical commercial or hospital...

Страница 25: ...rengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipm...

Страница 26: ...aximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 17 P 80 MHz to 800 MHz d 1 17 P 800 MHz to 2 5 GHz d 2 33 P 0 01 0 12 0 12 0 23 0...

Страница 27: ...27 DECLARATION OF CONFORMITY 12 DECLARATION OF CONFORMITY...

Страница 28: ...28 DECLARATION OF CONFORMITY PSPEC233 version 4 Terumo BCT Inc 10811 W Collins Ave Lakewood CO 80215 USA Registered trademark Copyright 2015 Terumo BCT Inc All rights reserved...

Отзывы: