20
CLEANING
9.2
Hand Unit
Clean once per week or more frequently, as required, if spillage of blood occurs.
For cleaning the electrodes remove the moving electrode.
Removal of moving electrode:
1.
Pull the blue trigger to “click” to release the moving electrode. Push spring lock
forward.
2.
First push inward, then pull out the electrode.
Clean the handle and both electrodes with a soft lint free cloth moistened with mild
detergent. Dry carefully and ensure that the electrodes are completely dry to prevent sparks.
To clean the electrode shaft, use a dry cotton swab.
After cleaning, inspect the electrodes for any mechanical damage or wear out. Do not use
damaged part.
For proper function, add a drip of light machine oil on electrode shaft after each cleaning!
Assemble the parts in reverse order.
Make sure the trigger and spring lock is in full forward position before electrode insertion.
Insert the moving electrode ensuring that it is positioned so as to be parallel with the fixed
electrode. Press the trigger twice and ensure that the moving electrode moves smoothly, is
in position and not loose.
CAUTION
The hand unit is not waterproof. Do not, under any circumstances,
submerge the hand unit in any kind of liquid. Doing so will generate
tiny electric discharges, cause the device to malfunction, and void the
warranty.
NOTES
Some sealing tests are recommended before resuming use.
Compare the pattern against picture in section 7.2.
Do not push the trigger in after removing the electrode since this may
block the locking mechanism.
If electrode cannot be inserted after cleaning, the trigger is probably
pressed during cleaning. To insert the front electrode again, see
chapter 8.2.