APPENDIX A-6: Terminal Command Designators
Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)
A-34
APPENDIX A-6: Terminal Command Designators
Table A-7: Terminal Command Designators
COMMAND
ADDITIONAL COMMAND SYNTAX
DESCRIPTION
? [ID]
Display help screen and commands list
LOGON [ID]
password
Establish connection to instrument
LOGOFF [ID]
Terminate connection to instrument
SET ALL|name|hexmask
Display test(s)
LIST [ALL|name|hexmask] [NAMES|HEX]
Print test(s) to screen
name
Print single test
T [ID]
CLEAR ALL|name|hexmask
Disable test(s)
SET ALL|name|hexmask
Display warning(s)
LIST [ALL|name|hexmask] [NAMES|HEX]
Print warning(s)
name
Clear single warning
W [ID]
CLEAR ALL|name|hexmask
Clear warning(s)
ZERO|LOWSPAN|SPAN [1|2]
Enter calibration mode
ASEQ number
Execute automatic sequence
COMPUTE ZERO|SPAN
Compute new slope/offset
EXIT
Exit calibration mode
C [ID]
ABORT
Abort calibration sequence
LIST
Print all I/O signals
name[=value]
Examine or set I/O signal
LIST NAMES
Print names of all diagnostic tests
ENTER name
Execute diagnostic test
EXIT
Exit diagnostic test
RESET [DATA] [CONFIG] [exitcode] Reset
instrument
PRINT ["name"] [SCRIPT]
Print DAS configuration
RECORDS ["name"]
Print number of DAS records
REPORT ["name"] [RECORDS=number] [FROM=<start
date>][TO=<end date>][VERBOSE|COMPACT|HEX]
(Print DAS records)(date format: MM/DD/YYYY(or YY)
[HH:MM:SS]
Print DAS records
D [ID]
CANCEL
Halt printing DAS records
LIST
Print setup variables
name[=value [warn_low [warn_high]]]
Modify variable
name="value"
Modify enumerated variable
CONFIG
Print instrument configuration
MAINT ON|OFF
Enter/exit maintenance mode
V [ID]
MODE
Print current instrument mode
DASBEGIN [<data channel definitions>] DASEND
Upload DAS configuration
CHANNELBEGIN propertylist CHANNELEND
Upload single DAS channel
CHANNELDELETE ["name"]
Delete DAS channels
The command syntax follows the command type, separated by a space character. Strings in [brackets] are optional
designators. The following key assignments also apply.
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...