Teledyne API – Model T300/T300M CO Analyzer
Setup Menu
103
5.4.3.2.
DUAL RANGE MODE (DUAL)
Selecting the
DUAL
range mode allows the
A1
and
A2
outputs to be configured with
different reporting ranges. The analyzer software calls these two ranges low and high.
•
The
LOW
range setting corresponds with the analog output labeled
A1
on the rear
panel of the instrument.
•
The
HIGH
range setting corresponds with the
A2
output.
While the software names these two ranges low and high, they do not have to be
configured that way. For example: The low range can be set for a span of 0-1000 ppm
while the high range is set for 0-500 ppm.
In
DUAL
range mode the
RANGE
test function displayed on the front panel will be
replaced by two separate functions:
•
RANGE1
: The range setting for the A1 output.
•
RANGE2
: The range setting for the A2 output.
To select the
DUAL
range mode press following buttonstroke sequence
.
SAMPLE
RANGE=50.0 PPM CO=XX.XX
<TST TST> CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE MODE MENU
MODE
SET UNIT
DIL
EXIT
SETUP X.X
RANGE MODE:SNGL
SNGL
DUAL
AUTO
ENTR EXIT
SETUP X.X
RANGE MODE:DUAL
SNGL DUAL AUTO
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT DIL
EXIT
When the instrument’s range mode is set to DUAL, the concentration field in the upper
right hand corner of the display alternates between displaying the low range value and
the high range value. The concentration currently being displayed is identified as
follows:
C1
=
LOW
(or
A1
) and
C2
=
HIGH
(or
A2
).
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
In DUAL range mode the LOW and HIGH ranges have separate slopes
and offsets for computing CO concentrations. The two ranges must be
independently calibrated.
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...