APPENDIX A-2: Setup Variables For Serial I/O
A-11
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
ET_TARGET_DET mV
4375
0–5000
Target detector reading during
electrical test.
ET_TARGET_CONC PPM
40,
400
3, 8
1–9999.99
Target concentration during
electrical test.
ET_CONC_RANGE Conc.
50,
5000
3, 8
0.1–50000
D/A concentration range during
electrical test.
STD_TEMP ºK
321
1–500
Standard temperature for
temperature compensation.
STD_PRESS "Hg
28.5,
28.7
8
,
28.8
12, 18
,
28.1
4
1–50
Standard pressure for pressure
compensation.
48
BENCH_SET ºC
Warnings:
43–53
0–100
Optical bench temperature set
point and warning limits.
68,
62
19,23
WHEEL_SET ºC
Warnings:
63–73,
57–67
19,23
0–100
Wheel temperature set point and
warning limits.
50
O2_CELL_SET
14
ºC
Warnings:
45–55
30–70
O
2
sensor cell temperature set
point and warning limits.
STD_O2_CELL_TEMP
14
ºK
323
1–500 Standard O
2
cell temperature for
temperature compensation.
ZERO_APPLY_IN_CAL
5
—
ON
OFF,
ON
ON
applies auto-reference offset
and dilution factor during
zero/span calibration;
OFF
disables both.
(Only applicable if
ZERO_ENABLE
is
ON
.)
ZERO_DWELL
3, 5, 8
Seconds,
Minutes
5
7,
3
5
1–60,
1–30
5
Dwell time after closing or
opening zero scrubber valve.
ZERO_SAMPLES
3, 5, 8
Samples
15,
750
5
,
1000
19
1–1000
Number of zero samples to
average.
ZERO_FILT_SIZE
3, 5, 8
Samples
5,
1
5
1–100
Auto-zero offset moving average
filter size.
ZERO_LIMIT
3, 5, 8
Ratio
1.2,
1.15
3, 8
,
1
5
0–5
Minimum auto-zero ratio allowed;
must be
greater
than this value
to be valid.
ZERO_CAL
3, 5, 8
Ratio
1.18
0.5–5
Calibrated auto-zero ratio.
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...