Teledyne API – Model T300/T300M CO Analyzer
Troubleshooting and Service
271
CO
2
Probe
Flow / Pressure
Sensor PCA
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
Sample Gas
Flow Control
INSTRUMENT CHASSIS
SAMPLE GAS
INLET
EXHAUST
GAS OUTLET
Particulate Filter
PUMP
GFC Wheel
Housing
GFC Motor Heat Sync
SAMPLE CHAMBER
Figure 12-12: T300/T300M – Internal Pneumatics with CO
2
Sensor Option 67A
12.2.2.
TYPICAL SAMPLE GAS FLOW PROBLEMS
12.2.2.1.
FLOW IS ZERO
The unit displays a SAMPLE FLOW warning message on the front panel display or the
SAMPLE FLOW test function reports a zero or very low flow rate.
Confirm that the sample pump is operating (turning). If not, use an AC voltmeter to
make sure that power is being supplied to the pump if no power is present at the
electrical leads of the pump.
1. If AC power is being supplied to the pump, but it is not turning, replace the pump.
2. If the pump is operating but the unit reports no gas flow, perform a flow check as
described in Section 11.3.4.
3. If no independent flow meter is available:
•
Disconnect the gas lines from both the sample inlet and the exhaust outlet on
the rear panel of the instrument.
•
Make sure that the unit is in basic SAMPLE Mode.
•
Place a finger over an Exhaust outlet on the rear panel of the instrument.
•
If gas is flowing through the analyzer, you will feel pulses of air being expelled
from the Exhaust outlet.
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...