Communications Setup and Operation
Teledyne API – Model T300/T300M CO Analyzer
162
6.7.3.
HESSEN PROTOCOL GAS LIST ENTRIES
6.7.3.1.
HESSEN PROTOCOL GAS ID
The T300/T300M Analyzer keeps a list of available gas types. Each entry in this list
takes the following format:
[GAS TYPE],[RANGE],[GAS ID],[REPORTED]
WHERE:
GAS TYPE
= The type of gas to be reported (e.g. CO, CO
2
, O
2
, etc.).
RANGE
= The concentration range for this entry in the gas list. This
feature permits the user to select which concentration range
will be used for this gas list entry. The T300/T300M Analyzer
has two ranges:
RANGE1
or LOW &
RANGE2
or HIGH (See
Section 5.4.1).
0 - The HESSEN protocol to use whatever range is currently
active.
1 - The HESSEN protocol will always use
RANGE1
for this
gas list entry.
2 - The HESSEN protocol will always use
RANGE2
for this
gas list entry.
3 - Not applicable to the T300/T300M Analyzer.
GAS ID
= An identification number assigned to a specific gas. In the
case of the T300/T300M Analyzer in its base configuration,
there is only one gas CO, and its default GAS ID is 310. This
ID number should not be modified.
REPORT
= States whether this list entry is to be reported or not reported
when ever this gas type or instrument is polled by the
HESSEN network. If the list entry is not to be reported this
field will be blank.
While the T300/T300M Analyzer is a single gas instrument that measures CO, it can
have additional, optional sensors for CO
2
or O
2
installed. The default gas list entries for
these three gases are:
CO, 0, 310, REPORTED
CO
2
, 0, 311, REPORTED
O
2
, 0, 312, REPORTED
These default settings cause the instrument to report the concentration value of the
currently active range. If you wish to have just concentration value stored for a specific
range, this list entry should be edited or additional entries should be added to the list.
EXAMPLE: Changing the above CO gas list entry to read:
CO, 2, 310, REPORTED
would cause only the last CO reading while
RANGE2
(HIGH) range was active to be
recorded.
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...