Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
201
9.2.2.
PERFORMING A BASIC MANUAL CALIBRATION CHECK
Wait until
STABIL
falls
below
0.2 PPM (for T300)
or 1.0 PPM (for T300M).
This may take several
minutes.
SAMPLE
STABIL= XXXX PPM
CO=XX.XX
< TST TST > CAL SETUP
The
ZERO
and/or
SPAN
keys will appear at various
points of this process.
It is not necessary to press
them.
SAMPLE RANGE=50.0PPM CO=XX.XX
< TST
TST >
CAL MSG SETUP
Set the Display to show
the
STABIL
test function.
This function calculates
the stability of the CO
measurement.
Toggle
TST>
button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls
below or 0.2 PPM (for
T300) or 1.0 PPM (for
T300M).
This may take several
minutes.
Allow span gas to enter the sample port
at the rear of the analyzer.
Analyzer display
continues to cycle
through all of the
available gas
measurements
throughout this
procedure.
Record CO zero point readings
Record CO span point readings\
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...