Section 8 Maintenance
8-19
with three small holes, are facing upwards towards the
outlet. See Figure 8-12.
Figure 8-12 Correct orientation of cartridge
2. Tighten check valve holders finger tight, then with pliers
tighten an additional ¼ turn.
8.7 Injection Valve Maintenance
Combi
Flash
Rf
+
systems, including those equipped with 4x,
have an automatic injection valve. Under normal operating
conditions, this self-cleaning valve does not require any
maintenance.
If maintenance is required, refer to Instruction Sheet
60-5232-475, Combi
Flash
Rf Injection Valve Replacement.
8.8 Electrical Troubleshooting
If your instrument stops working and the touch panel display
is off, check the line cord connection.
If the line cord is connected properly, check the fuses in the
power input module located on the system’s rear panel. These
are time delay fuses (T-type) and rated according to the mains
input power voltage.
Combiflash Rf
+
, Rf
+
Purion, and Rf 150 systems:
• 100 to 129 VAC mains input: 2.0 A fuse, one required
• 215 to 264 VAC mains input: 1.0 A fuse, two required
Combiflash Rf
+
Lumen system:
• 100 to 129 VAC mains input: 3.0 A fuse, one required
• 215 to 264 VAC mains input: 3.0 A fuse, two required
Содержание CombiFlash EZ Prep
Страница 2: ...CombiFlash Rf User Manual ii ...
Страница 8: ...CombiFlash Rf User Manual viii ...
Страница 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Страница 85: ...CombiFlash Rf User Manual 3 20 ...
Страница 103: ...CombiFlash Rf User Manual 4 18 ...
Страница 185: ...CombiFlash Rf User Manual 6 64 ...
Страница 195: ...CombiFlash Rf User Manual 7 10 Figure 7 4 Local Area Connection Properties Figure 7 5 Alternate Configuration Settings ...
Страница 199: ...CombiFlash Rf User Manual 7 14 ...
Страница 223: ...CombiFlash Rf User Manual 8 24 ...
Страница 225: ......
Страница 227: ......
Страница 229: ......
Страница 231: ......
Страница 233: ......
Страница 235: ......
Страница 237: ......