Section 1 Introduction
1-15
Electrical Safety per EN 61010-1
Pollution Degree
2
Installation Category
II
Maximum Altitude
2000 meters
Note 1. All specifications are subject to change.
Note 2. For columns undetected by RFID, the max pressure is limited to 50 psi.
Table 1-5 CombiFlash EZ Prep Specifications
a
Dimensions
67 cm (26.5")
43 cm (17")
43 cm (17") (H
W
D)
Weight
37.4 kg (82.5 lbs.),
43.6 kg (96.1 lbs.) with ELSD detector
Power Options
Input voltage range from 100 to 240 VAC
,
50/60 Hz,
300VA maximum typical.
Line cord is the disconnect device.
Line Frequency
50/60 Hz
Ambient Temperature
20 to 40 °C (maximum temperature must be at least
15 °C below the boiling point of the solvent)
Humidity (when
connected to power)
0 to 95% non-condensing
Flow Rate Range
5 to 200 mL/min
Flow Rate Accuracy
±5% (5-200 mL/min)
Pressure Limit
b
241
bar
(3500
psi)
Pressure Accuracy
5% of full scale
Gradient Formation
Binary gradient from solvent inlets 1-4. Solvent
selection can be automatically controlled by the
method.
Gradient Accuracy
±3% (20-200 mL/min)
Table 1-4 Combi
Flash
Rf+ 4x System
Specifications
1
(Continued)
Содержание CombiFlash EZ Prep
Страница 2: ...CombiFlash Rf User Manual ii ...
Страница 8: ...CombiFlash Rf User Manual viii ...
Страница 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Страница 85: ...CombiFlash Rf User Manual 3 20 ...
Страница 103: ...CombiFlash Rf User Manual 4 18 ...
Страница 185: ...CombiFlash Rf User Manual 6 64 ...
Страница 195: ...CombiFlash Rf User Manual 7 10 Figure 7 4 Local Area Connection Properties Figure 7 5 Alternate Configuration Settings ...
Страница 199: ...CombiFlash Rf User Manual 7 14 ...
Страница 223: ...CombiFlash Rf User Manual 8 24 ...
Страница 225: ......
Страница 227: ......
Страница 229: ......
Страница 231: ......
Страница 233: ......
Страница 235: ......
Страница 237: ......