CombiFlash® Rf
+
User Manual
1-14
Power Options
Input voltage range from 100 to 240 VAC, 50/60 Hz,
2.5 Amps system maximum, 0.5 amps module
maximum.
Line cord is the disconnect device.
Line Frequency
50/60 Hz
Ambient Temperature
20 to 40 °C (maximum temperature must be at least
15 °C below the boiling point of the solvent)
Humidity (when
connected to power)
95% relative humidity maximum at 20 to 40 °C
Flow Rate Range
1 to 200 mL/min
Flow Rate Accuracy
±5% (5-200 mL/min)
Pressure Limit
2
1379 kPa (200 psi)
Pressure Accuracy
5% of full scale
Gradient Formation
Binary gradient from A1, A2, and B1, B2 solvent inlets.
Solvent selection can be automatically controlled by
the method.
Peak Detection Modes
Slope or threshold
Flow Cell Path Length
0.1 mm, ±25%
UV Detection and
Monitor Wavelength
Selectable from 200 to 400 nm range. Optional 200 to
800 nm UV-Vis detector.
Wavelength Accuracy
±5 nm
Fraction Accuracy
±[2mL
(flow rate
60)]
Table 1-4 Combi
Flash
Rf+ 4x System
Specifications
1
(Continued)
Содержание CombiFlash EZ Prep
Страница 2: ...CombiFlash Rf User Manual ii ...
Страница 8: ...CombiFlash Rf User Manual viii ...
Страница 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Страница 85: ...CombiFlash Rf User Manual 3 20 ...
Страница 103: ...CombiFlash Rf User Manual 4 18 ...
Страница 185: ...CombiFlash Rf User Manual 6 64 ...
Страница 195: ...CombiFlash Rf User Manual 7 10 Figure 7 4 Local Area Connection Properties Figure 7 5 Alternate Configuration Settings ...
Страница 199: ...CombiFlash Rf User Manual 7 14 ...
Страница 223: ...CombiFlash Rf User Manual 8 24 ...
Страница 225: ......
Страница 227: ......
Страница 229: ......
Страница 231: ......
Страница 233: ......
Страница 235: ......
Страница 237: ......