3-1
Combi
Flash
®
Rf
+
User Manual
Section 3 Preparation:
Rf 150
Systems
This section provides instructions for unpacking and installing
the Combi
Flash
Rf 150 system. To prepare the system for
operation, sequentially follow all instructions in sections 3.1
through 3.11.
Note
Section 3.13 contains an Installation Qualification checklist. If
required, sign off the checklist entries as you successfully
complete the following sections.
Note
The Rf150 is rated for operation with reagent grade solvents.
Operation at pressures > 5 bar (~75 psi) with high purity hexanes
(>99% pure) at 100% concentration may cause the pump to stall
and produce a grating sound. Many times, normal operation can
be restored by removing the high pressure load and flushing the
pump with isopropyl alcohol.
3.1 Unpacking the Unit
The Combi
Flash
Rf 150 system is shipped in a single carton.
Carefully unpack the shipment and inspect the contents.
WARNING
The system is heavy. Use a two-person lift to prevent injury.
CAUTION
Do not lift the system by the fraction collector arm. Use the lifting
handles located on the instrument side panels.
Содержание CombiFlash EZ Prep
Страница 2: ...CombiFlash Rf User Manual ii ...
Страница 8: ...CombiFlash Rf User Manual viii ...
Страница 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Страница 85: ...CombiFlash Rf User Manual 3 20 ...
Страница 103: ...CombiFlash Rf User Manual 4 18 ...
Страница 185: ...CombiFlash Rf User Manual 6 64 ...
Страница 195: ...CombiFlash Rf User Manual 7 10 Figure 7 4 Local Area Connection Properties Figure 7 5 Alternate Configuration Settings ...
Страница 199: ...CombiFlash Rf User Manual 7 14 ...
Страница 223: ...CombiFlash Rf User Manual 8 24 ...
Страница 225: ......
Страница 227: ......
Страница 229: ......
Страница 231: ......
Страница 233: ......
Страница 235: ......
Страница 237: ......