Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
71
3.
Press the ENT key.
4.
In the “Peak Alarm Setting” item selection screen that
appears, point the cursor to any item you want to set by
pressing the UP or DOWN
key. Then press the ENT key.
5.
Enter numeric values by using the UP or DOWN keys.
Use the SIDE key to move the cursor to the right. Press
the ENT key to save the change.
To close the “Peak Alarm Setting” or cancel the mode midway,
press the ESC key. A previous screen will return.
Description of setting items:
Peak Alarm:
ON/OFF of peak alarm
Alarm Value:
If measuring value exceeds the set alarm value in
ppm, a peak counter increments by 1.
Alarm Count:
When a peak in excess of the alarm value occurs,
a peak count is incremented. The alarm count is
Содержание 7600
Страница 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Страница 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Страница 37: ...Infrared Gas Analyzer Operation Teledyne Analytical Instruments 37 Figure 3 3 Display Modes and Menu Hierarchy ...
Страница 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Страница 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Страница 119: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 119 A 8 Terminal Block Connections ...
Страница 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Страница 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Страница 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Страница 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Страница 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Страница 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...