Maintenance Model
7600
Teledyne Analytical Instruments
98
Figure 4-5: Detector Secure Nut
5.
Using the furnished cell mounting tool, turn the window
fixture to the left and remove it from the cell. See Figure
4-6.
6.
For cleaning the infrared transmission window and cell
inside surface, first eliminate coarse dust using a soft
brush and then wipe clean with a soft rag.
CAUTION:
THE WINDOW IS EASILY SCRATCHED. USE
EXTREME CARE WHEN PERFORMING THIS
OPERATION SO AS NOT TO DAMAGE IT.
7.
After cleaning, reinstall the sample cell and perform an
optical zero adjustment (see Section 4.4) and moisture
interference compensation adjustment (see Section 4.5).
Содержание 7600
Страница 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Страница 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Страница 37: ...Infrared Gas Analyzer Operation Teledyne Analytical Instruments 37 Figure 3 3 Display Modes and Menu Hierarchy ...
Страница 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Страница 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Страница 119: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 119 A 8 Terminal Block Connections ...
Страница 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Страница 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Страница 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Страница 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Страница 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Страница 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...