Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
57
Figure 3-6: Hysteresis Example for a High Limit Alarm
An alarm output is turned ON if measurement value exceeds the
upper limit value as shown in Figure 3-6. Once the alarm output has
been turned ON, it is not turned OFF until the process value falls below
the hysteresis width.
Note: The hysteresis setting is common to all alarms.
To set the hysteresis value:
1.
From the “Alarm No. Selection” screen, point the cursor
to “Hysteresis” by pressing the UP or DOWN key then
press ENT.
2.
In the “Hysteresis Value Setting” screen that appears,
enter hysteresis values using the UP or DOWN keys to
increment or decrement a digit. Use the SIDE key to
move to the next digit. When finished, press ENT.
Содержание 7600
Страница 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Страница 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Страница 37: ...Infrared Gas Analyzer Operation Teledyne Analytical Instruments 37 Figure 3 3 Display Modes and Menu Hierarchy ...
Страница 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Страница 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Страница 119: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 119 A 8 Terminal Block Connections ...
Страница 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Страница 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Страница 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Страница 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Страница 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Страница 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...