Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
61
Any operation other than “Stop Auto Calibration” (see
Section 3.5.4.5) is not permitted during auto calibration.
“Stop Auto Calibration” cannot be performed when the key
lock is ON.
To cancel auto calibration, set the key lock to OFF and
then execute “Stop Auto Calibration”.
Auto Calibration Example
Start Time
SUN
12:00
Cycle 1
day
Flow Time
Zero
350 sec
Ch1 Span
350 sec
Ch1 Span
350 sec
Ch1 Span
350 sec
Ch1 Span
300 sec
Ch1 Span
300 sec
EX. time
300 sec
ON/OFF ON
Figure 3-7 shows the auto calibration with the above settings.
Figure 3-7: Example Auto Calibration
Содержание 7600
Страница 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Страница 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Страница 37: ...Infrared Gas Analyzer Operation Teledyne Analytical Instruments 37 Figure 3 3 Display Modes and Menu Hierarchy ...
Страница 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Страница 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Страница 119: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 119 A 8 Terminal Block Connections ...
Страница 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Страница 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Страница 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Страница 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Страница 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Страница 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...