Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
87
1.
From the output adjustment screen which accessible from
the maintenance mode menu, move the cursor with the
UP or DOWN key to the output (OUT No. and zero/
span) to be adjusted, and then press the ENT key.
2.
The selected value is highlighted. Adjust the value using
the UP or DOWN key while watching the output on the
digital multi-meter. Press the SIDE key to select the next
digit.
3.
On completion of the adjustment, press the ENT key.
3.6.7 Other Parameters
The last submenu of the maintenance mode menu contains a group
of parameters that may be modified by the user. These include:
•
Password Set: Set the password used to enter the
maintenance mode. An arbitrary 4-digit
number can be selected.
•
O
2
Ref. Value: Set the oxygen concentration reference
value at the time of oxygen correction
calculation. Settable in the range from 00
to 19%.
•
Limit:
Set the oxygen concentration limit at the
time of oxygen correction calculation.
Settable in the range from 01 to 20%. See
O
2
correction concentration value in
Section 3.3.1
Measurement Mode
for
oxygen correction calculation procedure.
Содержание 7600
Страница 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Страница 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Страница 37: ...Infrared Gas Analyzer Operation Teledyne Analytical Instruments 37 Figure 3 3 Display Modes and Menu Hierarchy ...
Страница 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Страница 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Страница 119: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 119 A 8 Terminal Block Connections ...
Страница 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Страница 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Страница 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Страница 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Страница 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Страница 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...