Operation Model
7600
Teledyne Analytical Instruments
66
3.
In the “Setting of Auto Zero Calibration” screen that
appears, point the cursor to any item you want to set by
pressing the
UP or DOWN
key. Press the ENT
key.
4.
In the “Auto Zero Calibration Parameter Setting” screen
that appears, perform the value entry or the setting. For
the value entry or setting change, use the UP or DOWN
key. To change the setting, use the SIDE key to move the
cursor to the right.
5.
After changing the setting, press the ENT
key to accept
the change. Auto zero calibration will now be carried out
using the entered setting value.
To close the “Setting of Auto Calibration” or cancel the mode
midway, press the ESC key. A previous screen will return.
The contacts for the auto zero calibration are closed during auto
zero calibration.
Auto Zero Calibration Example
Start Time
SUN
12:00
Cycle 12
hour
Flow Time
300
sec
ON/OFF ON
Figure 3-8 shows the auto zero calibration with the above settings.
Содержание 7600
Страница 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Страница 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Страница 37: ...Infrared Gas Analyzer Operation Teledyne Analytical Instruments 37 Figure 3 3 Display Modes and Menu Hierarchy ...
Страница 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Страница 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Страница 119: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 119 A 8 Terminal Block Connections ...
Страница 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Страница 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Страница 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Страница 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Страница 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Страница 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...