Operation Model
7600
Teledyne Analytical Instruments
52
(AR), it will be calibrated in the range set in this menu when auto
calibration is performed. From this menu you can enable or disable the
auto calibration feature.
To navigate to the Auto Calibration Components/Range menu:
1.
From the measurement mode, press the MODE key to
display the user mode.
2.
Point the cursor to “Calibration Parameters” by pressing
the UP or DOWN key and press the ENT key.
3.
In the “Calibration Parameters” screen that appears, point
the cursor to “Auto Calibration Components/Range” by
pressing the UP or DOWN key. Press the ENT key.
4.
In the “Calibration Parameters” screen that appears, point
the cursor to “Auto Calibration Components / Range” by
pressing the UP or DOWN key. Then press the ENT key.
To set the range used for auto calibration:
1.
In the “Auto Calibration Components / Range” selection
screen that appears, point the cursor to the channel whose
Содержание 7600
Страница 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Страница 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Страница 37: ...Infrared Gas Analyzer Operation Teledyne Analytical Instruments 37 Figure 3 3 Display Modes and Menu Hierarchy ...
Страница 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Страница 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Страница 119: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 119 A 8 Terminal Block Connections ...
Страница 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Страница 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Страница 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Страница 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Страница 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Страница 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...