Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
59
5.
In the “Setting of Auto Calibration” screen that appears,
point the cursor to any item you want to set by pressing
the
UP or DOWN
key. Press the ENT key.
6.
In the “Auto Calibration Parameter Setting” screen that
appears, perform the value entry or the setting. For the
value entry or setting change, use the UP or DOWN key.
To change the setting, use the SIDE key to move the
cursor to the right.
7.
After changing the setting, press the ENT
key to accept
the change. Auto calibration will now be carried out
using the entered setting value.
To close the “Setting of Auto Calibration” or cancel the mode
midway, press the ESC key. A previous screen will return.
3.5.4.2
G
AS
F
LOW
T
IME
S
ETTING
In this submenu you select the gas whose flow time is to be
changed. The flow time range is from 60 to 900 seconds. The initial
value is set for 300 seconds.
Содержание 7600
Страница 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Страница 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Страница 37: ...Infrared Gas Analyzer Operation Teledyne Analytical Instruments 37 Figure 3 3 Display Modes and Menu Hierarchy ...
Страница 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Страница 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Страница 119: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 119 A 8 Terminal Block Connections ...
Страница 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Страница 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Страница 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Страница 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Страница 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Страница 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...