Teledyne API - T102/501 TRS, Addendum to T101 Operation Manual
TROUBLESHOOTING and REPAIR
71
11.5.1.
UV Lamp Adjustment and/or Replacement
There are three ways in which ambient conditions can affect the UV Lamp output and
therefore the accuracy of the TRS concentration measurement: lamp power supply failing,
lamp aging, and lamp positioning.
Lamp Power Supply Failing
If the UV lamp power supply begins to fail, it can cause the UV lamp value to decrease
in its output. Ensure the Lamp driver board has a Yellow Dot on the board. If it
doesn’t, contact TAPI Technical Support for a replacement board.
Lamp Aging
Over a period of months, the UV energy will show a downward trend, usually 30% -
50% in the first 90 days, and then a slower rate, until the end of useful life of the lamp.
Periodically running the UV lamp calibration routine (see the
Lamp Calibration
section
in the T101 Manual - PN 07266) will compensate for this until the lamp output becomes
too low to function at all, 2-3 years nominally.
Lamp Positioning
The UV output level of the lamp is not even across the entire length of the lamp. Some
portions of the lamp shine slightly more brightly than others. At the factory the position
of the UV lamp is adjusted to optimize the amount of UV light shining through the UV
filter/lens and into the reaction cell. Changes to the physical alignment of the lamp can
affect the analyzer’s ability to accurately measure SO
2
.
11.5.1.1.
Adjusting the UV Lamp
(
Peaking the Lamp
)
CAUTION:
ALWAYS wear UV-Protective, Safety Glasses when working with the
UV Lamp Assembly
1.
Set the analyzer display to show the signal I/O function,
UVLAMP_SIGNAL (see the Using the Diagnostic Signal I/O Function
section in the T101 Manual - PN 07266). UVLAMP_SIGNAL is function
35.
2.
Slightly loosen the large brass thumbscrew located on the shutter
housing (see Figure 11-1) so that the lamp can be moved.
3.
While watching the UVLAMP_SIGNAL reading, slowly rotate the lamp
or move it back and forth vertically until the UVLAMP_SIGNAL reading
is at its maximum.
NOTE:
DO NOT grasp the UV lamp by its cap when changing its position (see Figure 11-1).
Always grasp the main body of the lamp.
Ideally, the reading should be 3500mV±200mV.
•
If UVLAMP_SIGNAL is lower than 600mV, replace the lamp.
•
If UVLAMP_SIGNAL is greater than 3800 mV, adjust the pot on the
UV reference board down until the output reads 3500 mV, and then
continue to peak the lamp.
07267B DCN6485
Содержание 501 TRS
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Страница 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Страница 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Страница 54: ...07267B DCN6485 ...
Страница 90: ...TROUBLESHOOTING and REPAIR Teledyne API T102 501 TRS Addendum to T101 Operation Manual 88 07267B DCN6485 ...
Страница 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Страница 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Страница 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Страница 105: ...APPENDIX D Wire List and Electronic Schematics ...
Страница 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Страница 110: ...D 6 07267B DCN6485 ...
Страница 111: ......