Teledyne API - T102/501 TRS, Addendum to T101 Operation Manual
TROUBLESHOOTING and REPAIR
67
TEST
FUNCTION
NOMINAL
VALUE(S)
POSSIBLE CAUSE(S)
HVPS
≈
400 V to 900 V
Incorrect HVPS reading could be caused by; HVPS broken; preamp board circuit
problems.
RCELL TEMP
50ºC ± 1ºC
Incorrect temperature reading could be caused by: malfunctioning heater; relay
board communication (I
1
C bus); relay burnt out
BOX TEMP
ambient
+ ~ 5ºC
Incorrect temperature reading could be caused by: Environment out of
temperature operating range; broken thermistor; runaway heater
PMT TEMP
7ºC ± 2ºC
constant
Incorrect temperature reading could be caused by: TEC cooling circuit broken;
High chassis temperature; 12V power supply
IZS TEMP
(OPTION)
50ºC ± 1ºC
Malfunctioning heater; relay board communication (I
1
C bus); relay burnt out
PRESS
ambient
± 2 IN-HG-A
Incorrect SAMPLE pressure could be due to: pneumatic leak; malfunctioning
valve; malfunctioning pump; clogged flow orifices; sample inlet overpressure;
faulty pressure sensor
SLOPE
1.0 ± 0.3
Slope out of range could be due to: poor calibration quality; span gas
concentration incorrect; leaks; UV Lamp output decay.
OFFSET
< 250 mV
High offset could be due to: incorrect span gas concentration/contaminated zero
air/leak; low-level calibration off; light leak; aging UV filter; contaminated
reaction cell; pneumatic leak.
TIME OF DAY
Current time
Incorrect Time could be caused by: Internal clock drifting; move across time
zones; daylight savings time?
11.2.
M501-TRS Troubleshooting
11.2.1.
TRS Converter Not Heating:
Problems with heating the If the TRS converter oven can have several causes.
•
The “UUUU” error code shown on the M501-TRS temperature controller
display indicating a problem with the thermocouple
•
Check the resistance across the thermocouple leads for opens or
shorts.
•
Check to make sure that the thermocouple leads are securely
connected to the wiring block at the back of the controller.
•
Make sure that the wiring block/socket is correctly plugged onto
socket on the back of the controller.
•
The set point for the process value is set incorrectly.
•
Check the set value; if it is incorrect, reset it.
•
The controller is not in operational mode.
•
Press the PV/SV switch to return it to operation mode.
•
The heater is malfunctioning.
07267B DCN6485
Содержание 501 TRS
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Страница 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Страница 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Страница 54: ...07267B DCN6485 ...
Страница 90: ...TROUBLESHOOTING and REPAIR Teledyne API T102 501 TRS Addendum to T101 Operation Manual 88 07267B DCN6485 ...
Страница 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Страница 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Страница 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Страница 105: ...APPENDIX D Wire List and Electronic Schematics ...
Страница 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Страница 110: ...D 6 07267B DCN6485 ...
Страница 111: ......