Teledyne API - T102/501 TRS, Addendum to T101 Operation Manual
Pneumatic Connections
25
4.
PNEUMATIC CONNECTIONS
This section supersedes the information contained in the
Pneumatic Connections
section
in the T101 Manual - PN 07266.
NOTE
To prevent dust from getting into the analyzer, it was shipped with small plugs
inserted into each of the pneumatic fittings on the rear panel. Make sure that all dust
plugs are removed before attaching exhaust and supply gas lines.
CAUTION
Sample and calibration gases should only come into contact with PTFE
(Teflon) or glass materials. They should not come in contact with FEP or
stainless steel materials.
Figure 4-1 and Figure 4-2 show the most common configurations for gas supply and
exhaust lines to the Model T102 Analyzer. Figure 4-3, Figure 4-4, and Figure 4-5 show
the connections for units with valve options installed.
Please refer to Figure 3-4 and Figure 3-5 for the location of pneumatic connections at the
rear panel of the T102 and the M501-TRS instruments.
Table 4-1. Table 3-3: Inlet / Outlet Connector Labels and Functions
T102 PNEMATIC CONNECTERS
REAR PANEL LABEL
FUNCTION
SAMPLE
Connects the sample gas to the analyzer. When operating the analyzer without zero/span
option, this is also the inlet for any calibration gases.
EXHAUST
Exhausts the gas sampled by the analyzer. Connect to an outside area away from people.
SPAN
On units with zero/span/shutoff valve options installed, connect a gas line to the source of
calibrated span gas here.
ZERO AIR
On Units with zero/span valve or IZS option installed, this port connects the zero air gas or
the zero air cartridge to the analyzer.
TO CONVERTER
Sample gas leaves the T102 to be conditioned by the M501-TRS via this port.
FROM CONVERTER
Sample gas returns to the T102 after being conditioned by the M501-TRS via this port.
M501-TRS PNEUMATIC CONNECTERS
REAR PANEL LABEL
FUNCTION
FROM ANALYZER
Sample gas enters the M501-TRS from the T102 via this port.
TO ANALYZER
Sample gas leaves the M501-TRS to return to the T102 via this port.
07267B DCN6485
Содержание 501 TRS
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Страница 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Страница 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Страница 54: ...07267B DCN6485 ...
Страница 90: ...TROUBLESHOOTING and REPAIR Teledyne API T102 501 TRS Addendum to T101 Operation Manual 88 07267B DCN6485 ...
Страница 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Страница 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Страница 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Страница 105: ...APPENDIX D Wire List and Electronic Schematics ...
Страница 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Страница 110: ...D 6 07267B DCN6485 ...
Страница 111: ......