7.4.2.
Aborting the Autotune Process
1.
Press the DATA key once. The display will show
(1).
2.
Press the 1’s UP key once. The
will begin blinking
3.
Press the DOWN key once. The display will show
(zero).
4.
Press the ENTER key once.
5.
Press the PV/SV mode key to return to operational mode.
Note
Auto-tuning MUST to be repeated if there is a significant change in the set value.
If the temperature begins to oscillate excessively around the set value, it may be
necessary to repeat the autotune procedure.
7.5.
M501TRS Alarm Relay Adjustment
To set the High and Low Alarm points:
1.
Press the SELECT key once. The main display will show
2.
Use the SELECT , DOWN or 100’s UP key to scroll through the primary
menu parameters until the display shows either
(AL = Alarm Low)
or
(AH = High Alarm).
3.
Press the DATA key once. The current value of the alarm limit will be
displayed.
4.
To set each digit:
a.
Press the up-arrow under that digit once. The digit will flash.
b.
To increment that digit, press and hold the digit until the appropriate
number is displayed.
c.
To decrement that digit press and hold the DOWN key until the
appropriate number is displayed.
d.
To increment/decrement the 1000’s digit it is necessary to adjust
increment/decrement the 100’s digit up and down. Each time the
100’s digit passes “0” the 1000s digit will increment or decrement
correspondingly.
5.
Once the desired value is reached, press the ENT key to store the new
set value
6.
Press the PV/SV mode key to return to operational mode.
07267B DCN6485
Содержание 501 TRS
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Страница 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Страница 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Страница 54: ...07267B DCN6485 ...
Страница 90: ...TROUBLESHOOTING and REPAIR Teledyne API T102 501 TRS Addendum to T101 Operation Manual 88 07267B DCN6485 ...
Страница 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Страница 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Страница 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Страница 105: ...APPENDIX D Wire List and Electronic Schematics ...
Страница 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Страница 110: ...D 6 07267B DCN6485 ...
Страница 111: ......