Teledyne API - T102/501 TRS, Addendum to T101 Operation Manual
Pneumatic Connections
29
Figure 4-4. Pneumatic Connections for Formal Calibration of Units w/ IZS Valve Option
Source of
SAMPLE Gas
VENT if input is pressurized
Ambient
Air
MODEL
T102
From Converter
Sample
Zero Air (Scrubber)
Exhaust
Span
M501-TRS
To Converter
From Analyzer
To Analyzer
Figure 4-5. Pneumatic Connections for Informal Calibration Checks of Units with IZS
Valve Option
NOTE
Gas flow must be maintained at all times for units with IZS Options installed. The IZS
option requires a permeation tube (customer supplied) which emits H
2
S. Insufficient
gas flow can build up H
2
S to levels that will damage the instrument.
Remove the permeation device when taking the analyzer out of operation.
07267B DCN6485
Содержание 501 TRS
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Страница 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Страница 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Страница 54: ...07267B DCN6485 ...
Страница 90: ...TROUBLESHOOTING and REPAIR Teledyne API T102 501 TRS Addendum to T101 Operation Manual 88 07267B DCN6485 ...
Страница 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Страница 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Страница 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Страница 105: ...APPENDIX D Wire List and Electronic Schematics ...
Страница 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Страница 110: ...D 6 07267B DCN6485 ...
Страница 111: ......